Mirza Biscevic – Unipolar, Bipolar or Total Hip Endoprosthesis after Femoral Neck Fracture;

tra. 6, 2015

0

Mirza Biscevic – Unipolar, Bipolar or Total Hip Endoprosthesis after Femoral Neck Fracture;

Posted in : Mirza on by : AdminMirza
  • ,
  • ,
  • ,
  • ,
  • Original study

     

    Unipolar, Bipolar or Total Hip Endoprosthesis after Femoral Neck Fracture;

    What is the Right Decision?

     

    Dragica Smrke1, Mirza Biščević2.

     

    1Department for Traumatology, Clinical center Ljubljana, Slovenia,

    2Department for Orthopaedics and Traumatology, Clinical center Sarajevo, Bosnia and Herzegovina.

     

    – Department for Traumatology, Clinical center Ljubljana, Slovenia, and Department for Orthopaedics and Traumatology, Clinical center Sarajevo, Bosnia and Herzegovina.

     

    – Dr sci Mirza Biščević, dr. med. (corresponding author)

     

    – right endoprosthesis after hip fracture

    – no financial or any material support from any institution or corporation.

    – 1623 words, 9091 characters

    – no figures, one table.

    Unipolar, Bipolar or Total Hip Endoprosthesis after Femoral Neck Fracture;

    What is the Right Decision?

     

    Aim of this study was to recognize differences in long-term clinical outcome after femoral neck fracture and hip endoprosthesis implantation.

    Total of 145 patients were examined, 32 patients with unipolar, 70 with bipolar and 43 patients with total hip endoprosthesis.

    The mean values of Harris hip score, after 3.8±1.9 years, were: 72.1±17.8, 74.27±19.1, 78.2±22.5 for patients with unipolar, bipolar and total hip endoprosthesis, respectively. No statistically significant difference was observed (p=0.704). The in-hospital mortality rates were: 4.3%, 4.6%, and 5.3% for groups of patients with bipolar, unipolar and total hip endoprosthesis, respectively.

    Considering clinical outcomes, general health and costs, it could be concluded that choice of endoprosthesis does not pose an obstacle in patient’s recovery.

     

    Key words: hip, fracture, endoprosthesis, result, function.

     

     

    INTRODUCTION

    Consequences of aging on the hip joint (osteoporosis, varisation of the femoral neck), as well as reduction of psycho-physical abilities of a patient (neuromuscular incoordination, fear from activities) and comorbidities predispose elderly to the hip joint fracture. Cumulative risk was shown to be for women 18% and for men 6% (1). Approximately 30% of bed capacity in surgical facilities is occupied by patients with this sort of injury (2). The cost of treatment for patients with the hip fracture in the U.S. is exceeding $ 8.7 billions per year, and an assessment is that it would exceed 16 billions by the year 2040 (3).

    For all given reasons, a choice for the most rational treatment carries medical and economic importance for both patient and society. However, opinions about type of endoprosthesis that should be implanted after femoral neck fracture differ from one author to another. Some authors prefer implantation of unipolar partial hip endoprosthesis while others prefer bipolar partial endoprosthesis after displaced femoral neck fracture in elderly persons (4-19). Total hip endoprostheses, according to majority, achieve the best long-term clinical outcome. But its disadvantages, compared with partial endoprosthesis are: more extensive and longer operative procedures, increased blood loss, increased infection risk and mortality, longer rehabilitation period and higher costs (20, 21). Total hip endoprosthesis was shown to be a satisfactory salvage procedure after failure of other surgical solutions for femoral neck fracture (22).

    “In vitro” experiments cannot offer absolutely accurate data, due to many complex characteristics of structures of implant, bone, cement, and some uncertain numerical parameters. This emphasizes a need for clinical studies, which could ensure more realistic data about behavior of different endoprosthetic devices “in vivo” in order to confirm present methods in treatments and point to some critical moments. There is no published study which compared long-term clinical outcome among unipolar, bipolar and total hip endoprostheses after femoral neck fracture.

    Aim of this study was to compare long-term clinical outcomes after implantation of three different types of hip endoprostheses after femoral neck fracture.

     

    PATIENTS AND METHODS

    In this prospective study we collected hospital archive data only for patients with following inclusion criteria:

    – dislocated medial femoral neck fracture (Garden type III and IV),

    – absence of X ray visible degenerative pathological changes,

    – unilateral lesion,

    – implantation of cemented hip endoprosthesis as primary procedure by lateral surgical hip approach,

    – minimal follow up of 18 months,

    – surgical procedure performed in period Jan/88-Jun/01.

    The patients were divided in three groups:

    –     The first group consisted of 32 patients treated with partial unipolar hip endoprosthesis «Austin Moore, Instrumentaria DOO Zagreb, Croatia»,

    –     The second group consisted of 70 patients treated with partial bipolar hip endoprosthesis «Vario-cup, Link GmbH&Co Hamburg, Germany» and

    –     The third group consisted of 43 patients treated with total hip endoprosthesis «Lubinus, Link GmbH&Co Hamburg, Germany». Three groups differed only by the type of implanted endoprosthesis.

    The unipolar partial endoprosthesis is constructed as one piece; the head, the neck and the stem are made from the same material, and differ only in radius of the head and length of the neck and the stem. The bipolar endoprosthesis consists of a smaller polyethylene cup, tightened to the stem. The external cup`s outside surface articulates with the acetabulum and inside surface with the internal cup. This double mobility decreases number of motions between the acetabular cartilage and the articulation surface of the endoprosthesis, and consequently, acetabular erosion. In case of complications, it is possible to transform bipolar endoprosthesis into the total hip endoprosthesis by an implantation of an artificial acetabulum, while unipolar partial endoprosthesis has to be completely removed in case of certain complications.

    All patients were followed up for a mean period of 3.8 ± 1.9 years and their clinical outcomes were evaluated by Harris hip score.

    Harris hip score evaluates following hip parameters:

    hip pain (painless hip ………………………………..44 points),

    presence of deformity (no deformity …………………4 points),

    range of motions (full motions ………………………5 points),

    functional tests (complete functionality………………47 points):

    completely healthy hip…………….100 points.

    We have chosen Harris Hip Score for evaluation of long term clinical outcome because of its high validity and accuracy and its easy comparison with other evaluation scales (23, 24). Henning deems that satisfactory and better clinical outcomes are over 50 points of HHS, but for Lestrange, fair clinical outcome are values HHS over 70 points (25, 26).

    Statistical analysis was performed by Chi2 test and one-way ANOVA (Analysis of variance). The p value of 0.05 was deemed statistically significant.

     

    RESULTS

    Hospital archive data for 692 patients was collected according above mentioned inclusion criteria. Patients lost to follow up, dead patients and patients with incomplete data were excluded from the further analysis. In examined group of one hundred forty five patients, 85% were female and the mean age was 76 ± 5.1 y. The mean period from injury to surgery was 2.9 ± 1.3 days. The most common comorbidities were: cardiovascular diseases in 66% (96 patients), diabetes mellitus in 17% (25 patients), neurological diseases in 13% (19 patients) and pulmonary diseases in 10% (14 patients).

    There were no statistical significant differences between three described groups in parameters which could influence the long-term clinical outcome (age, follow up, comorbidities).

    Total hip endoprosthesis achieved the highest values of the Harris hip score, 78.23 ± 22.46, which has not differed significantly (p=0.704) from scores of unipolar and bipolar hip endoprostheses.

    Table 1. Age, follow up, and frequency of comorbidities with level of significance in three groups.

    PARAMETER UNIPOLAR P. BIPOLAR P. TOTAL P.        P
    AGE 1 75 78 73 0.091
    FOLLOW UP1 3.31 3.91 4.08 0.073
    CARDIO VASCULAR2 53 68 70 >0.1
    NEUROLOGICAL2 16 11 13 >0.1
    PULMONARY2 12 10 11 >0.1
    DIABETES MELLITUS2 16 17 16 >0.1
    PAIN1 37.7 37.9 36.87 >0.1
    LIMPING1 6.27 9.63 7.77 0.001
    HHS1 72.06 74.27 78.23 >0.1
    IN-HOSPITAL MORTALITY2 4.3 4.6 5.3 >0.1

     

    1– numerical parameters measured in years, statistical difference calculated by ANOVA

     

    2 – frequencies of diseases measured in percentages, statistical difference calculated by Chi2-test

     

    The mean of HHS values in all three groups (74.95 ± 19.52 96), can be considered as fair, according to Lestrange (26). The patients with implanted unipolar endoprosthesis had lower in-hospital mortality rate, but statistically not significant.

     

    DISCUSSION

    In our study, the majority of patients were octogenarian females with numerous comorbidities. Comorbidities and pre-injury conditions are significant factors that influence hip fracture after mild trauma (27). Since a percentage of elderly people is on rise in nowadays, geriatric diseases and injuries are becoming more important issue. Due to difficulties in treatment of the hip fracture, recovery is exhausting for patients and expensive for society.

    Harris hip score is an objective measure of long-term clinical outcome and the best tool in estimating treatment’s success. An average value of HHS 74.95 ± 19.52 (similar in other studies) indicated that a patient with implanted hip endoprosthesis after femoral neck fracture had periodic pain that did not affect her/his activities, could walk without problems at least 500 meters, limped to a certain degree, used a cane, could climb stairs holding a handrail, sat in the chair for a long time, put on shoes and socks with minor difficulties, used the public transportation and had no distinct deformity of the hip. With that level of hip functionality, our patients after endoprosthetic hip replacement are able for independent living, what is the main personal, medical and social goal for their age. Although average HHSs in all three groups were in domain of “fair” (by Lestrange), HHS of the bipolar partial endoprosthesis evaluated in this study is 2 points beyond HHS of the unipolar endoprosthesis (74 vs. 72), and only 4 points below the total endoprosthesis (74 vs. 78).

    The mean in-hospital mortality rate in all three groups (4.7%) was similar to Goldhill’s report (5.7%), and Lyons’s report (4.3%) (28,29). Slightly higher in-hospital mortality rate in group of patients with implanted total hip endoprosthesis (5.3%) compared with groups of patients with partial hip endoprostheses (4.5%) could be attributed to longer and more extensive surgery of implantation of total hip endoprosthesis and increased blood loss during surgery. Patients with implanted bipolar endoprosthesis had the lowest level of pain, limp and in-hospital mortality rate. Ichihashi supported use of the bipolar endoprosthesis in the femoral neck fracture and even after avascular necrosis of the femoral head, though he was uncertain about use of this endoprosthesis in patients who suffered from the hip arthrosis (30).

    Considering clinical results of this study, blood loss, duration of surgical procedure, possibility of revision, time of functional recovery and price of endoprosthesis, all types of endoprostheses are valuable for surgery of the hip. Although our groups were uniform and their HHS values were similar, decision about type of endoprosthesis should not be uniform. Total endoprosthesis is the logical choice in patients with previously damaged hip, unipolar endoprosthesis is the most rational choice for patients with short life expectancy. For others with displaced femoral neck fractures, bipolar partial endoprosthesis seems to be acceptable and compromising way of treatment.

    The limitations of this study are as follows: relatively small number of participants, the lack of severity of illness score and relatively short follow up. However, studies with similar limitations have been published in the literature and may have helped clinicians in decision making. Clearly, more comprehensive study on this subject is lacking and it may help to further answer new issues that arise.

    In conclusion, a choice of endoprosthesis is not the crucial obstacle in patient’s recovery to a pre-surgery state. It is only natural that a choice of an implant must be evaluated in accordance with clinical benefits of a chosen endoprosthesis on the one hand, and with general condition of a patient and economic factors on the other.

    LITERATURE

    1. Meunier PJ. Prevention of hip fractures. Am J Med. 1993 Nov 30;95(5A):75S-78S.
    2. Parker MJ, Myles JW, Anand JK. Cost-benefit analysis of hip fracture treatment. J Bone Joint Surg Br. 1992;74(2):261-4.
    3. Lavernia C, Lyon R. The short-term economic implications of prosthetic selection in hemiarthroplasty of the hip. Am J Orthop. 1998; 27 (6):415-3.
    4. Calder SJ, Anderson GH, Jagger C, Harper WM, Gregg PJ. Unipolar or bipolar prosthesis for displaced intracapsular hip fracture in octogenerians. J Bone Joint Surg Br.1996;78(B):391-4.
    5. Cornel CN, Levine D, O’ Doherty J, Lyden J. Uniporal versus bipolar arthroplasty for the treatment of femoral neck fractures in the elderly. Clin Orthop Relat Res.1998 Mar;(348):67-71.
    6. Raia FJ, Chapman CB, Herrera MF, Schweppe MW, Michelsen CB, Rosenwasser MP. Unipolar or bipolar hemiarthroplasty for femoral neck fractures in the elderly? Clin Orthop Relat Res. 2003 Sep;(414):259-65.
    7. Gilbert MS, Capozzi J. Unipolar or bipolar prosthesis for displaced intracapsular hip fractures?: An unaswered question. Clin Orthop Relat Res. 1998 Aug;(353):81-5.
    8. Hudson JI, Kenzora JE, Hebel JR, Gardner JF, Scherlis L, Epstein RS, Magaziner JS. Eight-year outcome associated with clinical options in the management of femoral neck fractures. Clin Orthop Relat Res. 1998 Mar;(348):59-7.
    9. Long JW, Knight W. Bateman UPF prosthesis in fractures of femoral neck. Clin Orthop Relat Res. 1980 Oct;(152):198-201.
    10. Ong BC, Maurer SG, Aharonoff GB, Zuckerman JD, Koval KJ. Unipolar versus bipolar hemiarthroplasty: functional outcome after femoral neck fracture at a minimum of thirty-six months of follow-up. J Orthop Trau. 2002;16(5):317-22.
    11. Overgaard S, Jensen TT, Bonde G, Mossing NB. The uncemented bipolar hemiarthroplasty for displaced femoral neck fractures. Acta Orthop Scand. 1991 Apr;62(2):115-20.
    12. Wathne RA, Koval KJ, Aharonoff GB, Zuckerman JD, Jones DA. Modular unipolar versus bipolar prosthesis: a prospective evaluation of functional outcome after femoral neck fracture. J Orthop Trauma. 1995; 9(4):298-4.
    13. Devas M, Hinves B. Prevention of acetabular erosion after hemiarthroplasty for fractured neck of femur. J Bone Joint Surg Br. 1983 Nov;65(5):548-51.
    14. Kenzora J. Outcome after hemiarthroplasty for femoral neck fractures in the elderly. Clin Orthop Relat Res. 1998 Mar;(348):51-8.
    15. Rodriguez-Merchan EC. Displaced intracapsular hip fractures: hemiarthroplasty or total arthroplasty? Clin Orthop Relat Res. 2002 Jun;(399):72-7.
    16. Nottage WM, McMaster WC. Comparison of bipolar implants with fixed-neck prosthesis in femoral neck fractures. Clin Orthop Relat Res. 1990 Feb;(251):38-43.
    17. Reymond MA, Kohler O, Chevalley JP, Rothenbuhler JM, Regazzoni P. Is implantation of a unipolar femoral endoprosthesis obsolete? Helv Chir Acta. 1992; 58(5):693-5.
    18. Dixon S, Bannister G. Cemented bipolar hemiarthroplasty for displaced intracapsular fracture in the mobile active elderly patient. Injury 2004; 35(2):152-6.
    19. Yamagata M, Chao EY, Ilstrup DM, Melton LJ 3rd, Coventry MB, Stauffer RN. Fixed-head and bipolar hip endoprostheses: a retrospective clinical and roentgenographic study. J Arthroplasty. 1987;2(4):327-41.
    20. Smrke D, Beden R, Stankoviski V. Bipolar versus total hip endoprosthesis: functional results. Arch. Orthop Trauma Surg. 2000; 120(5-6):259-61.
    21. Eyssel M, Schwenk W, Badke A, Krebs S, Stock W. Total endoprosthesis or dual head prosthesis in endoprosthetic management of femoral neck fractures? Unfallchirurg. 1994 Jul;97(7):347-52.
    22. van der Lugt JC, Dijkstra SD. Satisfactory results of Stanmore total hip arthroplasty after failed osteosynthesis of the femoral neck. Acta Orthop Belg. 2004 Feb;70(1):25-30.
    23. Soderman P, Malchau H. Is the Harris hip score system useful to study the outcome of total hip replacement? Clin Orthop Relat Res. 2001 Mar;(384):189-97
    24. Dai KR, Dong F, Wang YY. Comparative study of three different hip function evaluation systems. Zhonghua Wai Ke Za Zhi. 1994 Sep;32(9):535-8.
    25. Henning F, Hoepffner HJ, Muth A. Indications for bipolar prosthesis in femoral neck fractures. A retrospective study of the prognosis in geriatric patients with bipolar prosthesis with reference to the preoperative health status. 1991 Aug;94(8):409-16.
    26. Lestrange NR. Bipolar arthroplasty for 496 hip fractures. Clin Orthop Relat Res. 1990 Feb;(251):7-19.
    27. Pai VS, Arden D, Wilson N. Fractured neck of femur in the mobile independent elderly patient: should we treat with total hip replacement? J Orthop Surg (Hong Kong). 2003 Dec;11(2):123-8.
    28. Goldhill VB, Lyden JP, Cornell CN, Bochman RM. Bipolar hemiarthroplasty for fracture of the femoral neck. J Orthop Trauma. 1991;5(3):318-24.
    29. Lyons R. Clinical outcomes and treatment of hip fractures. Am J Med. 1997; 103(2A) 51-12.
    30. Ichihashi K, Imura S, Oomori H, Gesso H. Stress analysis on the acetabular side of bipolar hemiarthroplasty by the two-dimensional finite element method incorporating the boundary friction layer. Nippon Seikeigeka Gakkai Zasshi. 1994 Nov;68(11):939-52.